Novo: NH6000sc Ammonium Analyser, 1 channel, includes integrated sample supply pump for FX610 filtration unit, flow detection, 230 VAC

NH6000sc Ammonium Analyser, 1 channel, includes integrated sample supply pump for FX610 filtration unit, flow detection, 230 VAC
Proizvod #: LXV461.99.01211
Jedinična cijena Kontaktirajte nas
Otprema unutar 1 tjedna

Maximise uptime and accuracy with trusted technology

The NH6000sc ammonium analyser from Hach® uses fast, accurate and trusted GSE technology. The NH6000sc provides reliable measurements with automatic calibration, validation, and cleaning. A verification grab sample improves consistency between online and lab data through an automatic sampling process. Predictive diagnostics and supportive maintenance workflows help you maximise uptime.

Save time with biannual maintenance

The NH6000sc ammonium analyser reduced routine user maintenance frequency to only twice per year. Engineered with efficiency in mind, the instrument's streamlined design and optimised reagents make operation simple and user-friendly. With a standard cleaning cycle, the reagents have an extended replacement time of every 6-months.

Robust and lightweight intergrated filtration options

The integrated FX610 filtration system and the more rugged FX620 feature lightweight construction with improved air cleaning and innovative flow detection. These features ensure reliable functionality and reduced operational touchpoints. Plus, you’ll have the ability to remotely check that your sample is consistently supplied to your analyser to increase confidence in your results.

Specifikacije

Br. kanala: 1
Certificiranja: CE, UKCA, CMIM, FCC, ISED, certified to UL and CSA safety standards by TÜV
Čuvanje uređaja: -20 - 60 °C; 95% relative humidity, non-condensing
Dimenzije: 575 x 991 x 425 mm (22.63 x 39.01 x 16.73 in.)
Duljina kabla: Data and power cable: 2 m from edge of enclosure
Granica detekcije: Measuring range 1: 0.02 mg/L NH₄-N
Measuring range 2: 0.05 mg/L NH₄-N
Measuring range 3: 1 mg/L NH₄-N
Measuring range 4: 10 mg/L NH₄-N
Izlaz : Relay, analog outputs, network interface through SC Controller
Klasa kućišta: Rating: IP55, NEMA 3R UL50E
Material: PUR 66
Klasa zaštite: Class 1
Kompatibilnost kontrolera: Digital SC Controller. Controllers are sold separately.
Kvaliteta uzorka: Ultra filtrated or comparable (FX6x0 < 0.45 µm)
Masa: Approximately 45 kg without chemicals
Materijal: PUR 66 UV-resistant
Mjerna metoda: Gas sensing electrode (GSE)
Mjerni interval: 5, 10, 15, 20 or 30 min. (user adjustable)
Mjerni rang: 0.02 - 5.0 mg/L NH₄-N
  0.05 - 20 mg/L NH₄-N
  10 - 1000 mg/L NH₄-N
  1 - 100 mg/L NH₄-N
Opcije/Mogućnosti: Flow detection
Parametar: Ammonium
Pauza: Power supplied by the SC controller. With the SC4500, an additional power box is needed.
Analyser and heated drain tubing (powered by analyser): 115 VAC or 230 VAC
pH rang: 5 - 9 pH
Ponovljivost: With Standard Solution:
Measuring range 1: 3% of the measured value + 0.02 mg/L
Measuring range 2: 2% of the measured value + 0.05 mg/L
Measuring range 3: 2% of the measured value + 1.0 mg/L
Measuring range 4: 2% of the measured value + 10 mg/L
Posebne napomene: The liquid level in the basin must be below the bottom of the analyser
Protok uzorka: 1.0 - 20.0 L/h to the analyser from the external sample filtration
Radna temperatura: -20 - 45 °C; 95% relative humidity, non-condensing
Stupanj zagađenosti: 2
Temperatura uzorka: 4 - 40 °C
Točnost: With Standard Solution:
Measuring range 1: ≤ 1 mg/L: 3% of the measured value + 0.02 mg/L > 1 mg/L: 5% of the measured value + 0.02 mg/L
Measuring range 2: 3% of the measured value + 0.05 mg/L
Measuring range 3: 3% of the measured value + 1.0 mg/L
Measuring range 4: 4.5% of the measured value + 10 mg/L
Tolerancija klorida: ≤ 1000 mg/L Cl-
Uključen sustav za filtraciju: Integrated sample supply pump for FX610 Filtration System
Unutarnji/Vanjski: Indoor or outdoor use
Uzorak: tlak: 2.5 bar (0.25 mPa) maximum
Visina: 2000 m maximum
Vrijeme odziva T90: Measuring range 1: ≤ 0.2 mg/L: 3 measurements required (15 total minutes minimum); > 0.2 mg/L: 1 measurement required (5 total minutes)
Measuring range 2: < 5 minutes
Measuring range 3: < 5 minutes
Measuring range 4: < 5 minutes