Alarma:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analogni izlazi:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Automatsko čišćenje:
Da
Broj uzoraka:
1, 2, 4, or 8
Certificiranja:
CE compliant / ETL certified
Demineralised water:
Consumption: 140 mL/analysis (hour)
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimenzije (V x Š x D):
690 mm x 465 mm x 330 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garancija:
24 mjeseci
Granica detekcije:
TN: ≤ 0.25 mg/L
TP: ≤ 0.01 mg/L
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencija :
TN: The main interferences are Br- and I-. When the amount of I- is 2.2 fold of the amount of TN, or the amount of Br- is 3.4 fold of the amount of TN, this will interfere on the test results. Dissolved organic matter, surfactants and Chromium (VI) interfere. Various inorganic substances not normally found in natural water, such as Chlorite [ClO₂-] and Chlorate [ClO₃-], may interfere.
TP: Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, and Vanadium (V), Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Kalibracija:
Automatic, 2-point; frequency freely programmable
Kivete:
Standard measurement cycle time for both TN & TP: 60 minutes
Klasa zaštite:
Analyser cabinet: IP55 / Panel PC: IP65
Kvaliteta uzorka:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Masa:
25 kg
Materijal:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Mjerna metoda:
TN: UV photometric measurement at 220 nm after persulphate digestion in alkaline medium, based on APHA 4500-NO3 (B)
TP: Colorimetric measurement at 700 nm using ascorbic acid reduction and molybdate colour solution after persulphate digestion in acidic medium, based on APHA 4500-P
Mjerni rang:
0.25 - 10 mg/L TN; 0.010 - 2 mg/L TP
Napajanje:
230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Odvod:
Atmospheric pressure, vented, min. Ø 32 mm
Parametar:
Total Nitrogen (TN), Total Phosphorus (TP)
Precision:
Better than 3% (TN) and 2% (TP) full scale range for standard test solutions
Protok uzorka:
100 - 300 mL/min
Reagent Requirements:
Keep between 10 - 30 °C
Što je u pretincu:
EZ7642 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)
Temperatura prostora:
10 - 30 °C, ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura uzorka:
10 - 30 °C
Uzorak: tlak:
By external overflow vessel
Validation:
Automatic; frequency freely programmable