Alarma:
1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analogni izlazi:
Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Automatsko čišćenje:
Da
Broj uzoraka:
1, 2, or 4
Certificiranja:
CE compliant / ETL certified
Cooling water:
Not required
Demineralised water:
For rinsing
Digital outputs:
Optional: Modbus (TCP/IP, RS485)
Dimenzije (V x Š x D):
690 mm x 465 mm x 330 mm
Earth connection:
Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garancija:
24 mjeseci
Granica detekcije:
≤ 1 µg/L
Instrument air:
Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencija :
Iodide, organic matter, copper Cu(II) > 30 µg/L, iron Fe(III) >20 mg/L, various metals in mg/L levels may interfere. Fats, oil, proteins, surfactants and tar.
Kalibracija:
Automatic, 2-point; frequency freely programmable
Kivete:
10 minutes
Klasa zaštite:
Analyser cabinet: IP44 / Panel PC: IP65
Kvaliteta uzorka:
Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Masa:
25 kg
Materijal:
Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Mjerna metoda:
Stripping voltammetry using gold electrode
Mjerni rang:
1 - 20 µg/L As
1 - 20 µg/L As
Model:
EZ6001
Napajanje:
100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Odvod:
Atmospheric pressure, vented, min. Ø 32 mm
Parametar:
Arsenic, total dissolved As(III+V)
Precision:
Better than 5% full scale range for standard test solutions
Protok uzorka:
100 - 300 mL/min
Reagent Requirements:
Keep between 10 - 30 °C
Što je u pretincu:
EZ6001 Arsenic Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x Reference Electrode (Ag/AgCl), 1 x Working Electrode - Gold, 1 Counter Electrode, 1 x empty 5L Reagent Container with Fittings (Buffer Solution) and 1 x empty 10L Reagent Container with Fittings (H2SO4) and 2 x empty 1L Reagent Containers with Fittings (Reference Blank & Reference As Solution)
Temperatura prostora:
10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura uzorka:
10 - 30 °C
Uzorak: tlak:
By external overflow vessel
Validation:
Automatic; frequency freely programmable