Specifikacije

Alarma: 1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analogni izlazi: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Broj uzoraka: 1, 2, 4 or 8
Certificiranja: CE compliant / UL certified
Demineralised water: For rinsing / dilution
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimenzije (V x Š x D): 690 mm x 465 mm x 330 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garancija: 24 mjeseci
Granica detekcije: ≤ 10 mg/L
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencija : Soaps, oily matter, suspended solids or precipitates may coat the glass electrode and cause a sluggish response. Allow additional time between titrant additions to let the electrode come to equilibrium or clean the electrodes occasionally.
Kalibracija: Automatic, 2-point; frequency freely programmable
Kivete: 10 - 15 minutes
Klasa zaštite: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Kvaliteta uzorka: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Masa: 25 kg
Materijal: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Mjerna metoda: Acid-base titration with hydrochloric acid
Mjerni rang: 100 - 5000 mg/L CaCO3

Optional:
10 - 500 mg/L CaCO3
25 - 1250 mg/L CaCO3
50 - 2500 mg/L CaCO3
Internal dispenser dilution (max. factor 100)
  100 - 5000 mg/L CaCO₃
Napajanje: 100 - 240 VAC, 50/60 Hz

Max. power consumption: 120 VA
Naziv metode: Titrimetric
Odvod: Atmospheric pressure, vented, min. Ø 32 mm
Parametar: Alkalinity, free
Precision: Better than 2% full scale range for standard test solutions
Protok uzorka: 100 - 300 mL/min
Reagent Requirements: Keep between 10 - 30 °C
Temperatura prostora: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura uzorka: 10 - 30 °C
Uzorak: tlak: By external overflow vessel
Validation: Automatic; frequency freely programmable