EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, 4x mA

EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, 4x mA
Proizvod #: EZ3005.9800140T
Jedinična cijena Kontaktirajte nas

Reliable, Stable Measurements

Avoid downtime from unreliable ISE data caused by temperature fluctuations. The EZ3005sc stabilises both sample and electrode using a temperature controlled flow cell, eliminating measurement errors and delivering precise, repeatable chloride results operators can trust.

Lower Maintenance Effort

Minimise downtime and service costs with the EZ3005sc. Automated 2 point calibration, reagent alerts, a rugged ISE electrode, and minimal moving parts extend maintenance intervals; so operators spend less time servicing instruments and more time on plant operations.

Reduced Operating Cost

The EZ3005sc helps optimise operating costs by reducing chemical consumption and service visits. Multi stream capability monitors up to eight sample lines, lowering cost per sampling point and extending equipment life.

Designed for easy integration into existing control systems

With a wide range of analogue and digital outputs, this analyser aligns with plant specific requirements, ensuring compatibility with existing infrastructure and immediate access to real time measurement data.

Što je u pretincu

EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)

Specifikacije

Analogni izlazi: Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
Automatsko čišćenje: Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Broj uzoraka: Standard: 1
Optional accessory for 2, 4, or 8 sample streams
Certificiranja: CE, ETL certified to UL and CSA safety standards, UKCA
Demineralised water: Does not apply
Digital outputs: Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150)

Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP

RS485 communication: Profibus DP or Modbus RTU
Dimenzije (V x Š x D): 688 mm x 460 mm x 340 mm
Earth connection: Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garancija: 24 mjeseci
Granica detekcije: 2 - 100 mg/L Clˉ: 2 mg/L
5 - 250 mg/L Clˉ: 5 mg/L
10 - 500 mg/L Clˉ: 10 mg/L
20 - 1000 mg/L Clˉ: 20 mg/L
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
Interferencija : Bromide [(Br)-], sulphide [(S)²-], iodide [(I)-], cyanide [(CN)-] ions may interfere. Mercury [(Hg)+] must be absent. Ammonia [NH₃] and thiosulphate [(S₂O₃)²-] may interfere. Fats, oil, proteins, surfactants and tar.
Kalibracija: Automatic; 2-point; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
Kivete: Default: 5 minutes
Continuous: 5 minutes
Klasa zaštite: IP44
Kvaliteta uzorka: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Masa: Max. 35 kg
Materijal: Hinged part: Thermoform ABS, Door: PMMA
Wall section: Galvanised steel, powder coated
Mjerna metoda: Discontinuous, direct measurement by combined Ion-Selective Electrode
  Ion-Selective Electrode
Mjerni rang: 2 - 100 mg/L Clˉ
5 - 250 mg/L Clˉ
10 - 500 mg/L Clˉ
20 - 1000 mg/L Clˉ
Napajanje: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Odvod: Atmospheric pressure, vented, min. Ø 32 mm
Parametar: Chloride
Precision: Better than 2% full scale range for standard test solutions
Protok uzorka: 100 - 300 mL/min
Reagent Requirements: Keep between 10 - 30 °C
Što je u pretincu: EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyser, 1 stream, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)
Temperatura prostora: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura uzorka: 10 - 30 °C
Uzorak: tlak: By external overflow vessel (open to atmospheric pressure)
Validation: Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly