EZ1003 Ammonium Analyser (Nessler Method), 1 stream, Modbus RS485

EZ1003 Ammonium Analyser (Nessler Method), 1 stream, Modbus RS485
Proizvod #: EZ1003.99001C02
HRK Cijena: Kontaktirajte nas

Specifikacije

Alarma: 1x malfunctioning, 4x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Broj uzoraka: 1 stream

Optional:
1 to 8 streams
Certificiranja: CE compliant / UL certified
Digital outputs: Modbus RS485
Dimenzije (V x Š x D): 690 mm x 465 mm x 330 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garancija: 2 godine
Granica detekcije: ≤ 250 µg/L
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferencija : Acetone, alcohols, aldehydes, aliphatic and aromatic amines, chlorine, glycine, organic chloramines and sulphide. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Izlaz : Modbus RS485

Optional:
Active 4 - 20 mA max. 500 Ohm load, 1 to 8 outputs
RS232, Modbus TCP/IP
Kalibracija: Automatic, 2-point; frequency freely programmable
Kivete: 6 min (dilution + 5 min)
Klasa zaštite: Analyser cabinet: IP55 / Panel PC: IP65
Kvaliteta uzorka: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Masa: 25 kg
Materijal: Hinged part: Thermoform ABS, door: plexiglass
Wall section: Galvanised steel, powder coated
Mjerna metoda: Colorimetric measurement at 450 nm conform with standard method EPA 350.1 (Nessler)
Mjerni rang: 0.25 - 2 mg/L NH4-N

Optional:
1 - 8 mg/L NH4-N (with internal dilution)
2 - 16 mg/L NH4-N (with internal dilution)
2.5 - 200 mg/L NH4-N (with internal dilution)
Napajanje: 110 - 220 VAC, 2 A, 50/60Hz
Max. power consumption: 150 VA
Other voltages available on request
Odvod: Atmospheric pressure, vented, min. Ø 64 mm
Parametar: Ammonium
Precision: Better than 2% full scale range for standard test solutions
Protok uzorka: 100 - 300 mL/min
Reagent Requirements: Keep between 10 - 30 °C
Temperatura prostora: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Temperatura uzorka: 10 - 30 °C
Uzorak: tlak: By external overflow vessel
Validation: Automatic; frequency freely programmable